2 General safety information
8
2

General safety information

DANGER
Potentially lethal shock hazard:
➢ It is essential that you switch the line voltage OFF at the main switch
of the building installation before beginning installation of the unit!
DANGER
Potentially lethal shock hazard!
Fixed connection!
Installing a mains plug instead of the specified fixed connection infringes
international medical regulatory actions and is prohibited. In case of
error, this puts patients, users, and other parties seriously at risk.
WARNING
Radiation protection
The valid radiation protection regulations and measures must be
observed. The statutory radiation protection equipment must be used.
In case of malfunctions, cancel the exposure immediately by letting go
of the exposure release button.
CAUTION
Risk of injury!
➢ The installation must be carried out in accordance with the
requirements stated in our Installation Instructions.
➢ Installation may be carried out only by personnel specifically
authorized by Dentsply Sirona.
CAUTION
Electromagnetic compatibility:
The unit should not be operated in the immediate vicinity of other
devices (see also "Installation Requirements, Chapter on EMC").
➢ If this proves to be unavoidable, the unit should be monitored to
ensure that it is operating properly.
NOTICE
Electrical components of the unit can be destroyed.
Prior to opening the unit
➢ Please comply with the usual precautionary measures for handling
printed circuit boards (ESD).
➢ Make sure you touch a ground point to discharge yourself prior to
touching the components.
➢ Use an ESD wrist band and connect it to the protective ground wire.
Dentsply Sirona
Installation Instructions Heliodent Plus
62 14 972 D 3507
D 3507.031.01.15.02 03.2018