Preparation for Use

Electrostatic discharge (ESD), excessive input signals, or mechanical shock
CAUTION
can dramatically degrade the performance of the probe. Be sure to observe
the following precautions.
– Never touch the tip of the probe!
– Always hold the probe by the retracted metal sleeve.
– Work at a workstation equipped with an anti-static mat.
– Extend the protective sleeve when not actually taking measurements.
– Wear an anti-static wrist strap and avoid introducing static electricity into
– Make sure the device under test (DUT) is at the same ground potential as

Power requirements

If using the probe with an instrument that does not supply probe power, you can
purchase an 85024A-001 adapter cable assembly and one of the following
Keysight external power supplies:
– E3620A
– E3630A
– E3631A
You need to ensure the correct setting of the power supply (+15 V and –12.6 V)
per attached labels on the 85024A-001 adapter cable assembly.
Keysight 85024A User's and Service Guide
the device under test (DUT) or test setup.
the probe.
Installation
3
29