Service

Cable/Probe wand replacement

Tools required:
– 10 mm open-end wrench
– 9/16 inch open-end wrench
– 1/8 inch-wide flatblade screwdriver
– Phillips No. 0 or positive screwdriver
– Long-nose pliers
Special tool:
– Spanner/wrench (part number 85024-20041)?
Refer to
1 Remove the nut with a 10 mm open-end wrench.
2 Remove the nose assembly and set aside.
The amplifier microcircuit is very sensitive to static electricity. Exercise full
CAUTION
anti-static precautions and use great care when performing the procedure.
Refer to
3 Remove the amplifier microcircuit. Remove the spring clip and elastic
4 Remove the regulator housing covers by performing the procedure in
Refer to
5 Remove the connector body with the 9/16 inch wrench.
70