7
Service
Isopropyl alcohol is extremely flammable, causes irritation, may cause eye
WARNING
damage, and is harmful if swallowed or inhaled. It may be harmful if
absorbed through the skin. Keep away from heat, sparks, and flame. Avoid
contact with eyes, skin, clothing. Avoid breathing vapor. Keep in tightly
closed container. Use with adequate ventilation. Wash thoroughly after
handling. In case of fire, use alcohol foam, dry chemical, or carbon dioxide:
water may be ineffective. In case of spill, soak up with sand or earth. Flush
spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state,
and local environmental regulations.
3 Clean the mating plane surfaces
4 Dry the connector
5 Reinspect
76
Apply a small amount of isopropyl alcohol to a new swab and clean the center
and outer conductor mating plane surfaces. When cleaning a female
connector, avoid snagging the swab on the center conductor contact fingers
by using short strokes.
After cleaning, blow the connector dry with a gentle stream of clean
compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
Inspect the connector again to make sure that no particles or residue are
present.