Performance Tests

Average Noise Level

Specifications (at 25 °C ±5 °C)
Description
This test procedure uses a power meter to measure the power level of the probe
output when the probe tip is connected through 50 ohms to ground.
Procedure
1 Connect the power sensor to the power meter and allow the power meter to
2 Calibrate and zero the power meter.
3 Enable averaging on the power meter.
4 Connect the power cable on the RF probe to a power source, allow the probe
5 Connect the probe tip adapter to the tip of the probe. (This effectively grounds
6 Connect the RF probe to the power sensor using a type-N barrel.
7 Read the power meter. If the reading is unstable, visually determine the
46
warm up for five minutes.
to warm up for at least three minutes.
the probe tip).
approximate peak value over a period of about twenty seconds. Record that
value on the test record.
<17 mV RTI (or –23 dBm at probe output)