10:1 Divider; Operating Characteristics; Inspection - Keysight 85024A User's And Service Manual

High frequency probe
Table of Contents

Advertisement

10:1 Divider

The 10:1 divider fits over the tip of the probe and provides the following changes
to the probe's operating parameters:
– Increases (by a factor of 10) the input voltage at which 1 dB compression
– Decreases the input capacitance without changing input resistance, thereby
Two or more dividers may be cascaded to provide higher divide ratios.

Operating characteristics

Divider ratio
Input capacitance
Input resistance
Input voltage for 1 dB compression 3.0 volts peak
Maximum rated DC input voltage
Maximum rated RF input voltage
Frequency range
[a] When used with the active probe.

Inspection

Make sure the tip is not bent or discolored. Periodically inspect the barrel of the
probe receptacle, making sure it is clean and free of grit. Clean the receptacle with
clean compressed air.
When cascading 10:1 dividers, periodically inspect and clean the exterior of the
metal sleeve.
Keysight 85024A User's and Service Guide
occurs.
decreasing capacitive loading.
10:1
Typically <0.7 pF
1 MW
[a]
±50 volts
15 volts peak
Same as the active probe (300 kHz to 3 GHz)