Performance Tests
14 Connect the power sensor to the output port of the analyzer. Adjust the
0.0 dBm is used to allow for possible inaccuracy of power meters
NOTE
15 Disconnect the power sensor from the output port of the analyzer. Connect the
16 Connect the probe, the probe adapter, and a barrel between the attenuator
17 Remove the attenuator pad, connect the probe adapter to the output port of
18 Perform the calculation indicated on the test record to determine the result of
44
analyzer power output level to produce a reading of 0.0 dBm on the power
meter.
attenuator to the output port of the analyzer.
and the power sensor. The signal from the analyzer should now go through the
attenuator and then the probe before reaching the power sensor. Record the
power reading on the
the analyzer. Connect the attenuator pad between the probe and the power
sensor. The signal from the analyzer should now go through the probe and
then the pad before reaching the power sensor. Record the power reading on
the
"Test Record"
on page 47 (item G).
the test.
"Test Record"
on page 47 (item F).
Keysight 85024A User's and Service Guide