Compatibility Between Bpl And Bpl Ii Batteries; Existing Bpl Batteries Are Compatible With Bpl Ii Handpieces - Synthes Battery Power Line II Instructions For Use Manual

Battery driven power tool system for orthopedics and traumatology
Hide thumbs Also See for Battery Power Line II:
Table of Contents

Advertisement

Introduction
Drive Units
Compatibility between BPL and BPL II batteries
Existing BPL handpieces are compatible with BPL II
batteries
The existing BPL handpieces (530.605, 530.610,
530.615) can be used with the new BPL II battery
(530.630), battery casing (530.690) and sterile cover
(530.660) as seen in Fig. 1.
Existing BPL batteries are compatible with BPL II
handpieces
The existing BPL battery (530.620), battery casing
(530.680) and sterile cover (530.650) can be used with
the new BPL II handpieces (530.705, 530.710, 530.715)
as seen in Fig. 2.
Note:
All BPL/BPL II attachments are fully
compatible with the BPL/BPL II handpieces
(530.605/530.705).
Note:
Technical data is subject to tolerances and
may vary when combining the two systems.
12
DePuy Synthes Battery Power Line II Surgical Technique
Figure 1
Figure 2
530.605
530.690
530.705
530.680