3. Spray and wipe
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes (Fig. 2). Follow the
enzymatic detergent manufacturer's directions for cor-
rect temperature, water quality (i.e. pH, hardness) and
concentration/dilution.

4. Rinse with tap water

Rinse device with cold tap water for a minimum of
2 minutes. Use a syringe or pipette to flush lumens and
channels.

5. Clean with detergent

Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris (Figs. 3 and 4). Follow
the enzymatic cleaner or detergent manufacturer's in-
structions for use for correct temperature, water quality
and concentration/dilution.
Note:
For the Quick Coupling for Pins B 3.2 mm
(530.796) the brush should only be inserted from the
front.
Figure 2
Figure 3
Figure 4: Quick Coupling for Kirschner Wires and for Pins B 1.5–4.0 mm
(530.791)