Follow the information in the "Care and Maintenance"
section starting on page 45 as well as the Instructions
for Use of the Synthes Universal Battery Charger II
(DSEM/PWT/1114/0050).
Precautions:
• The battery must never be washed, rinsed or
dropped. This would destroy the battery with
possible secondary damage.
• Generally, medical power tools will heat up if in
constant use. The cool down times should be
observed, see "Duty Cycle" section on page 76, in
order to prevent the power tool from exceeding its
acceptable surface temperature.
• In case of cell leakage, do not allow the leaking
fluid to come into contact with skin or eyes. In
case of contact, wash the affected area with copi-
ous amounts of water and seek medical advice.
• Faulty batteries must not be reused and should be
disposed of in an environmentally friendly man-
ner and in accordance with national regulations.
• Please refer to the regulations for transporting
Li-Ion batteries when returning them to the
Synthes Service Center.
Warnings:
• Risk of fire, explosion and burns. Do not disassem-
ble, crush, heat above 60°C/140°F or incinerate the
battery cells.
• Never expose batteries to temperatures over
60°C/140°F. The maximum exposure time at
60°C/140°F is 72 hours.
• Do not dismantle, open or shred batteries.
Battery Power Line II Surgical Technique DePuy Synthes
21