Battery Reamer/Drill II
Problem
Attachments cannot couple to drive
unit
Difficulty removing attachments
from drive unit
Bone, cutting tool and drive unit
heat up during surgery
Possible causes
Coupling is blocked by residue
Attachment coupling is damaged
Coupling is blocked by residue
Drive unit coupling sleeve is dam-
aged
The cutting tool is blunt
Battery Power Line II Surgical Technique DePuy Synthes
Solution
Precaution:
Immediately turn
mode switch OFF (Lock position).
Remove solid particles with pick-
ups. Clean and lubricate accord-
ing to care and maintenance
guidelines.
Send damaged attachment to
Synthes Service Center
Precaution:
Immediately turn
mode switch OFF (Lock position).
Remove solid particles with pick-
ups. Clean and lubricate coupling
sleeve according to care and
maintenance guidelines.
Send damaged drive unit to Synthes
Service Center
Replace the tool