Remove saw blade
Open the saw blade coupling fully by twisting the lock-
ing knob and remove the oscillating saw blade (Fig. 3).
Instructions for handling saw blades
Synthes recommends using a new saw blade for each
operation to ensure that the saw blade is optimally
sharpened and clean.
The following risks are associated with used saw blades:
• Thermal necrosis caused by excessive heat build-up
• Infection caused by residues
• Extended cutting time due to poor sawing perfor-
mance
• Potentially, splintering of the teeth or the saw blade
Figure 3
The use of irrigation fl uid is recommended to cool the
cutting tools and prevent heat necrosis.
Check the cutting tools for wear and/or damage after
each use, and replace if necessary. For optimal perfor-
mance only use Synthes saw blades. These are optimized
to meet the specifi c requirements of the tool. Non
Synthes saw blades can considerably reduce the lifetime
of the system.
Detailed ordering information on saw blades for the
Battery Power Line II system can be found in the
brochure "Saw Blades" (DSEM/PWT/0514/0004).