Removing and disassembling the battery pack

Press both release buttons simultaneously on the drive
unit to remove the battery pack (Fig. 9).
Open the casing by pressing both battery casing locks
and remove the battery or hold open the battery casing
to allow another person to remove the battery (Fig. 10).
Ensure that the battery does not touch the exterior of
the battery casing in order to avoid contaminating the
battery. If this occurs follow the information in the "Care
and Maintenance" section starting on page 45.
Store battery in Universal Battery Charger II (05.001.204)
when not in use (Fig. 11).
Alternatively please follow the guidelines provided in the
STERRAD/V-PRO sterilization guide
(DSEM/PWT/0615/0068). No other sterilization methods
are allowed.
Precautions:
Do not wash, rinse, drop or apply force
to the battery (530.630). This will destroy it with
possible secondary damage.
Figure 9
Figure 10
Figure 11
Battery Power Line II Surgical Technique DePuy Synthes
Circulating
person
Scrubbed person
19