2.10 XORB Board (PRODIGY I Systems DF+11999 and
lower only)
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-68 System Overview
The XORB printed circuit board provides protection to various circuits (low
voltage) from transients within the HVPS and tube head (see 2.1.1 for
location).
• There may be occasional static discharges within the Tube Head. The
transients caused by these static discharges (arcs) are shunted to
ground through the array of transorbs present on the XORB board.
• There is no safety hazard to personnel, however, the electronics could be
damaged were no protection provided.
PRODIGY Service Manual (Rev C - 2000)