wtw PurCon IS sample preparation Operating Manual page 24

Prepares the sample for the analysis of component substances with analyzers
Table of Contents

Advertisement

Commissioning
Configuring the
temperature monitoring
72
With the aid of a jumper in the temperature measuring box you can set
the operation mode of the temperature monitoring as follows:
ON
Jumper
OFF
AUTO
AUTO
ON
OFF
AUTO
AUTO
ON
OFF
AUTO
AUTO
AUTO (factory setting):
Automatic control of the heating tape.
The heating tape is switched on if the mea-
sured temperature falls below 5 °C.
The heating tape is switched off if the mea-
sured temperature rises above 8 °C.
Each switching is delayed by approx.
30 seconds.
OFF: The heating tape is always switched off
ON: The heating tape is always switched on
PurCon IS
ba43133e06
05/2009