Table 202: Guidelines And Manufacturer's Declaration Electromagnetic Interference - ergoline SUNRISE 480 Turbo Power Operating Instructions Manual

Table of Contents

Advertisement

Interference
immunity tests
Discharging static electricity
(ESD)
IEC 61000-4-2
Fast transient electrical noise /
bursts
IEC 61000-4-4
Surges
61000-4-5
Magnetic field radiated by supply
frequency (50/60 Hz)
IEC 61000-4-8
Voltage drops,
Short interruptions
and supply voltage
fluctuations
IEC 61000-4-11
06/2018
1019683-00
16.1.2
The Tanning system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
IEC 60601 -
Test level
±6 kV contact discharge
±8 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
3 A/m
<5% U
(>95% drop in U
T
T
for 0.5 periods
40% U
(60% drop in U
)
T
T
for 5 periods
70% U
(30% drop in U
)
T
T
for 25 periods
<5% U
(>95% drop in U
T
T
for 5 s
Note: U
is the mains alternating voltage used for the test levels.
T
Table 202: Guidelines and manufacturer's
declaration – Electromagnetic interference
Defined
level
±6 kV contact discharge
±8 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
3 A/m
)
<5% U
(>95% drop in U
T
for 0.5 periods
40% U
(60% drop in U
T
for 5 periods
70% U
(30% drop in U
T
for 25 periods
)
<5% U
(>95% drop in U
T
for 5 s
Appendix
Electromagnetic
environment –
Guidelines
)
T
)
T
)
T
)
T
Operating instructions – 73/84

Hide quick links:

Advertisement

Table of Contents
loading