Table 202: Guidelines And Manufacturer's Declaration Electromagnetic Interference - ergoline Beauty Angel 7200 Operating Instructions Manual

Hide thumbs Also See for Beauty Angel 7200:
Table of Contents

Advertisement

Appendix
Interference
immunity tests
Discharging static electricity
(ESD) IEC 61000-4-2
Fast transient electrical noise /
bursts IEC 61000-4-4
Surges
61000-4-5
Magnetic field radiated by
supply frequency (50/60 Hz)
IEC 61000-4-8
Voltage drops,
Short interruptions
and supply voltage
IEC 61000-4-11
80/92
16.1.2
The beauty system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
IEC 60601 -
Test level
±8 kV contact discharge
±2 kV air discharge
±4 kV air discharge
±8 kV air discharge
±15 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
30 A/m
<5% U
(>95% drop in U
T
for 5 s
Only this test is performed,
because the device used 2 or 3
phases and has more than 16A /
phase.
Note: U
is the mains alternating voltage used for the test levels.
T
Table 202: Guidelines and manufacturer's
declaration – Electromagnetic interference
Defined
level
±8 kV contact discharge
±2 kV air discharge
±4 kV air discharge
±8 kV air discharge
±15 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
No test performed. The device
does not contain magnetically
sensitive components or circuits.
)
<5% U
(>95% drop in U
T
T
for 5 s
Electromagnetic
environment –
Guidelines
)
T

Hide quick links:

Advertisement

Table of Contents
loading