Additional Information - GE MAC VU360 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Cleaning and Disinfection

Additional Information

• GE Healthcare used Super Sani-Cloth
validation. This cleaning agent is not listed in preference to other available
cleaning agents which may perform satisfactorily.
• The cleaning and disinfection information is provided in accordance with ANSI/
AAMIST81, ISO 17664. The recommendations provided above have been validated
as capable of preparing non-sterile GE Healthcare MAC VU360
It remains the responsibility of the user to make sure that the cleaning and
disinfection are performed using appropriate equipment, materials, and personnel
and achieves the desired result. This requires validation and routine monitoring
of the process. Any deviation from the provided recommendations should be
properly evaluated for effectiveness and potential adverse consequences.
• All users should be qualified personnel with documented expertise, competency,
and training. Users should be trained on hospital policies and procedures along
with current applicable guidelines and standards.
• Users should utilize appropriate Personal Protective Equipment (PPE) when
cleaning and disinfecting devices in accordance with the Department of
Environmental and Occupational Health and Safety's (OSHA) blood-borne
pathogen guidelines or equivalent.
120
  MAC VU360
Resting ECG Analysis System
®
wipes during cleaning and disinfection
medical devices.
 2088531-370-2

Advertisement

Table of Contents
loading

Table of Contents