June 2003
Making Electrical
Connections (cont.)
Installation Instructions
for CE Mark Conformity
1-2
Make connections by placing the press lock lever into the desired
terminal. One press lock lever is supplied with each terminal block.
Press and hold the lever against the terminal block and insert the
stripped and tinned portion of the wire into the terminal. Release the
lever to secure the connection.
TURN OFF THE SERIES 1
BEFORE MAKING ANY CONNECTIONS.
Proper connections and cabling are extremely important to accurate
measurement. Be sure to use the correct cable type for each probe and
make sure that cables are not damaged during installation. If you are
not using a GE Panametrics-supplied cable or are using a modified
cable, read the following section carefully.
ATTENTION EUROPEAN CUSTOMERS!
IN ORDER TO MEET CE MARK REQUIREMENTS,
YOU MUST INSTALL ELECTRICAL CABLES AS
DESCRIBED BELOW.
CE MARK COMPLIANCE IS REQUIRED ONLY FOR
UNITS USED IN EEC COUNTRIES.
For CE Mark compliance, you must shield and ground the electrical
connections as shown in Table 1-1.
Note: If you make the modifications as discussed in this table, your
unit will comply with the EMC Directive 89/336/EEC.
Table 1-1: Wiring Modifications for CE Compliance
Connection
Input/Output
Connect the shields to the nearest chassis
ground using the shortest run of wire possible.
After you make all the necessary electrical connections, seal the
unused cable entry holes with standard conduit plugs or equivalent.
!WARNING!
NOTICE
Termination Modification
Installing Optional Features