Calibration

Process calibration

Calibration establishes reference points that relate the detector output to actual (or known)
values of the process. When you first receive the scale there is no calibration and the sensor
does not read the weight.
There are three different types of calibrations:
1. Channel #1 Initial Cal—Perform an initial calibration with a static load to get a close
calibration (this is not an easily reproducible calibration)
2. Channel #1 Periodic Cal—This is a quick and reproducible calibration. It always ends
with the same calibration, without moving the product around
3. Channel #2 Cal.

Calibration main menus

All of the calibration functions for the gauge are available under the Cal Menu.
018 CAL MENU 1 OF 2
SET PRODUCT CODE
CALIBRATE GAGE
SET TIME CONSTANT

Product code screen

003 PRODUCT CODE
SELECTION SCREEN
OLD=0.0
EXAMINE P.C. TABLE
The product code groups together a set of parameters that are associated with a unique
product. There are nine Product codes available for use.
Product codes 0 through 8 are available for Channel #1 that handles the process sensor
input, and product code 9 is for Channel #2 that is for optional sensor input.
You can view Product Code Tables from access level 0, however, access level 1 enables you
to setup or change values in the table. The table in Appendix 1 lists the set of 25 parameters.
Procedures for changing the product code screen are available on page 66 of this manual.
82

Figure 37: Cal menu screens

NEW=0.0

Figure 38: Product code screens

W-4510 Weigh Scale Technical Reference Manual
018 CAL MENU 2 OF 2
TOTALIZER MENU
527 PR CODE
0
ITEM 000
NEXT=0.00
CURRENT VAL
NEW VAL
00000
00000