Step 1: Simple Cal On Low; Figure 43: Simple Re-Cal On Low Screen - Ohmart Vega W-4510 Technical Reference Hardware Manual

Weigh scale with gen 2000 electronics model / with frequency output
Table of Contents

Advertisement

Step 1: Simple cal on low

1.
From screen 149 CAL CHANNEL #1, select SIMPLE RE-CAL LOW and press
ENTER
2.
From screen 141 START SIMPLE CAL ON EMPTY, select YES and press
ENTER to confirm
3.
On screen 173 CAL TIME LEFT, press ENTER when the counter reaches 0
4.
From screen 174 EVAL DATA COLLECT, verify that the absorber and process
value are correct. If not correct, press SELECT for NO to adjust the CAL or YES to
continue. Press ENTER
5.
If you chose YES, screen 175 displays
6.
Connect the write protect jumper, JP17 and save to EEPROM
7.
On 175 SAVE CALIBRATE RESULT IN EEPROM screen, press SELECT to
choose YES or NO and press ENTER
8.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
W-4510 Weigh Scale Technical Reference Manual
141 START SIMPLE
CAL ON EMPTY
CONFIRM
YES
NO (ABORT)

Figure 43: Simple re-cal on low screen

Procedure 30: Simple re-cal Channel #1
Calibration
95

Advertisement

Table of Contents
loading

Table of Contents