Calibration
Step 6: Setting the absorber value
1.
With a moving empty belt, pull out the absorber and perform a simple data collect
to determine the value
Make a note of the absorber reading and counts
2.
From 159 ENTER ABSORBER VALUES screen, select HIGH ABSORBER
VALUE and press ENTER
3.
From 148 HIGH ABSORBER screen, enter the new value and press ENTER
4.
Screen 148 displays, press ENTER on the YES line to save to upper RAM
5.
Connect the write protect jumper, JP17 and save to EEPROM
6.
Disconnect the write protect jumper, JP17.
92
Procedure 28: Set absorber value
W-4510 Weigh Scale Technical Reference Manual