Appendix I: Parameter blocks
Screen 525 Totalizer Block
The base address for screen 525 TOTALS CHAN #1 is 529, item 1072.
Item#
Parameter Name
0
STATUS
1
DATA POINTER
2
REFERENCE
3
PRESET LOW ENTRY
4
PRESET HIGH ENTRY
5
REMOTE TOTALIZER
DRIVE
6
PRESET ALARM
7
LOW TOTAL
8
HIGH TOTAL
Item#
Parameter Name
8
STATUS
9
DATA POINTER
10
REFERENCE
11
PRESET LOW ENTRY
12
PRESET HIGH ENTRY
13
REMOTE TOTALIZER
DRIVE
14
PRESET ALARM
15
LOW TOTAL
16
HIGH TOTAL
134
525 TOTALS CHAN #1
ITEM
000
NEXT=000
CURRENT VAL
NEW VAL
00000
00000
Figure 67: Totalizer screen
Table 29: Screen 525 Totalizer block #1
0=normal counting, 1=Reset, 2=Hold, 3=Reset/Hold
Location of Rate data to totalize
Time and weight conversion from Rate to Total
Low 4-digits of 8-digit Preset entry
High 4-digits of 8-digit Preset entry
Changes to 1 for each count added into Total. Changes back to 0 in
50milliseconds. 10 counts per second maximum
Equals 1 when 8-digit Total exceeds 8-digit Preset Entry. Raise Preset or
Reset Total to clear alarm
Lower 4-digits of 8-digit totalizer
Upper 4-digits of 8-digit totalizer
Table 30: Screen 525 Totalizer block #2
0=normal counting, 1=Reset, 2=Hold, 3=Reset/Hold
Location of Rate data to totalize
Time and weight conversion from Rate to Total
Low 4-digits of 8-digit Preset entry
High 4-digits of 8-digit Preset entry
Changes to 1 for each count added into Total. Changes back to 0 in
50 milliseconds. 10 counts per second maximum
Equals 1 when 8-digit Total exceeds 8-digit Preset Entry. Raise Preset or
Reset Total to clear alarm
Lower 4-digits of 8-digit totalizer
Upper 4-digits of 8-digit totalizer
W-4510 Weigh Scale Technical Reference Manual
Parameter description
Parameter description