Calibration
Step 2: Cal Hi on absorber
1.
From screen 181 CAL Hi screen, select YES and press ENTER to confirm
2.
From screen 170 CAL TIME LEFT screen, press ENTER after timeout
3.
On screen 171 EVAL DATA COLL., view the absorber value and measured value.
Press ENTER on the YES field to adjust the cal or press SELECT and press
ENTER to choose NO
4.
Connect the write protect jumper, JP17 and save to EEPROM
5.
On 172 SAVE CALIBRATE RESULT IN EEPROM screen, press SELECT to
choose YES or NO and press ENTER
6.
Disconnect the write protect jumper, JP17.
96
181 CAL HI, ABSORBER
CONFIRM
YES
Figure 44: Cal on absorber screen
Procedure 31: Cal Hi on absorber—Channel #1
W-4510 Weigh Scale Technical Reference Manual
NO (ABORT)