Upper Ram Transfers - Ohmart Vega W-4510 Technical Reference Hardware Manual

Weigh scale with gen 2000 electronics model / with frequency output
Table of Contents

Advertisement

Setup

Upper Ram transfers

The upper RAM can hold a copy of the active gauge configuration. This is useful when
making a copy of the non-volatile EEPROM. Use this screen to make transfers to and from
upper RAM. Enter a one to initiate the transfer. The transfer completes almost
instantaneously.
1.
From 310 MEMORY BACKUP screen, select UPPER RAM TRANSFERS and
press ENTER
2.
From 314 UPPER RAM XFERS screen, select RAM TO UPPER RAM and press
ENTER
3.
In 316 RAM TO UPPER RAM TRANSFER screen, press 1 and press ENTER to
activate the transfer
1.
From 310 MEMORY BACKUP screen, select UPPER RAM TRANSFERS and
press ENTER
2.
From 314 UPPER RAM XFERS screen, press SELECT to choose UPPER RAM
TO RAM and press ENTER
3.
In 318 UPPER RAM TO RAM TRANSFER screen, press 1 and press ENTER to
activate the transfer
44
Procedure 2: Transfer to upper RAM
Procedure 3: Transfer to RAM
W-4510 Weigh Scale Technical Reference Manual