Selecting The Filter Type; Selecting A Filter Type, Damping, And Fast Cutoff - Ohmart Vega W-4510 Technical Reference Hardware Manual

Weigh scale with gen 2000 electronics model / with frequency output
Table of Contents

Advertisement

Selecting the filter type

Select the filter type at Item No. 18 in the Product Code Table on screen 527 (see the table,
on page 136, in Appendix I).
Note
Rectangular window is the default filtering type.

Selecting a filter type, damping, and fast cutoff

Procedure 17: Selecting a filter type, damping, and fast cutoff
1.
From 527 PR CODE screen, enter one of the following for a filter type (item 18):
0 for RC exponential
1 for rectangular window
Press ENTER
2.
From 018 CAL MENU 1 OF 2, press SELECT to choose SET TIME
CONSTANT and press ENTER
3.
From 035 TIME CONSTANT screen, enter the new value for the time constant
and press ENTER
4.
From 035 TIME CONSTANT screen, select ENTER FAST CUTOFF VALUE
and press ENTER
5.
From 049 ENTER DIGITAL FILTER CUTOFF VALUE screen, select
CHANNEL #1 FILTER and press ENTER
6.
From 050 ENTER CHANNEL#1 FILTER CUTOFF VALUE, enter the new value
and press ENTER
7.
Press PREVIOUS SCREEN to return to screen 049
8.
From 049 ENTER DIGITAL FILTER CUTOFF VALUE screen, select
CHANNEL #2 FILTER and press ENTER
9.
From 051 ENTER CHANNEL#2 FILTER CUTOFF VALUE screen, enter the
new value and press ENTER
10.
Note: To turn off Fast cutoff, enter 0 as the value.
17.
Connect write protect jumper, JP17 and save to EEPROM
18.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
W-4510 Weigh Scale Technical Reference Manual
Setup
75