Calibration
Step 4: Absorber value setup
1.
From screen 159 ENTER ABSORBER VALUES, select HI ABSORBER VALUE
and press ENTER
2.
From screen 148 HIGH ABSORBER screen, enter the new value of the absorber
on high and press ENTER and then press ENTER again to confirm
3.
Connect the write protect jumper, JP17 and save to EEPROM
4.
Disconnect the write protect jumper, JP17.
Step 5: Saving the calibration information to EEPROM
After completion of step 4, all of the calibration information is in the RAM memory of the
Smart Pro. You must copy this data to the non-volatile EEPROM for long-term storage. On all
system re-boots, the EEPROM data copies into the RAM memory and becomes the current
configuration. Refer to page 46 for information concerning the upload and download
configuration program. Follow the procedure on page 93 to save the calibration information to
the EEPROM.
98
159 ENTER ABSORBER
VALUES.
LOW ABSORBER VALUE
HI ABSORBER VALUE
Figure 46: Absorber value setup screen
Procedure 33: Absorber value setup—Channel #1
W-4510 Weigh Scale Technical Reference Manual