16.
Press PREVIOUS SCREEN to return to the MAIN MENU
17.
On the Main Menu screen, press select to position the cursor in the top left corner
and type 316 and press ENTER
18.
From 316 RAM TO UPPER RAM TRANSFER screen, press 1 to activate the
transfer
19.
Screen 316 displays, type 003 in the menu number area and press ENTER
20.
From 003 PRODUCT CODE SELECTION SCREEN, type 0 and press ENTER to
return to product code 0
21.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.

Measurement units setup

The measurement units for both the channels are set at screen 009. Under screen 9, the
weight and rate units list separately for Channel #1. The units for the second channel are
limited to speed and custom.

Custom units

If the engineering units required are not in the standard list, you can program a custom unit.
Please contact Ohmart for your specific needs.
W-4510 Weigh Scale Technical Reference Manual
Procedure 10:Setup gauge span—Channel #2 (continued)
009 SELECT UNITS
SELECT CHAN#1 UNITS
SELECT CHAN#2 UNITS

Figure 27: Measurement units setup screen

Setup
57