LED summary table
LED
Description
+6V
+6V DC voltage level
to electronics
+24V
Not used
Relay
Not used
LED
Description
Mem
Not used
HART
Not used
CPU
Central processing
unit on CPU board
"heartbeat"
Aux
Auxiliary input
frequency signal
indicator
HV
Not used
Field
Radiation field
indicator
W-4510 Weigh Scale Technical Reference Manual
Table 21: Power supply board LED summary table
Normal condition
Error condition
ON
OFF—Electronics are not
receiving +6VDC voltage
required for functioning
Table 22: CPU board LED summary table
Normal condition
Error condition
N/A
N/A
ON
N/A
Blinks at rate of
LED does not blink. CPU
1time/sec.
not functioning.
Blinks if auxiliary
None
input present. OFF—
if no auxiliary input
present.
N/A
N/A
Cycles in proportion
None
to radiation field
intensity at detector.
ON for 10sec. for
each mR/hr, then off
for two seconds.
(Can use LED 5 that
blinks 1time/sec to
time LED 9 for field
indicator.)
Diagnostics and repair
Recommendation
Verify +6V on test points.
Check fuse on Power Supply
board. Check power input
terminals 1, 2.
Recommendation
Check power input. Replace
CPU board.
Check auxiliary input wiring
terminals 11 & 12 with a meter
for frequency signal. Check
auxiliary input equipment.
A 1mR/hr (2,580nC/kg/hr) field
is usually required for a
measurement. Check for closed
source shutter, buildup, and
insulation.
123