Step 5: Calculating the calibration

The data collected in the previous section does not immediately impact the calibration
parameters the gauge is using. Once you finish collecting the data for both ends of the
calibration, calculate the new calibration parameters by using the "Two Point Cal" function
from the calibration menu.
Note: It is important to have the proper Linearization curve loaded into the gauge before
executing the Two Point Cal function. The gauge may have to extrapolate the endpoints and
must have the curve in place to do this.

Two point calibration

90
155 TWO POINT CAL
HAVE YOU DONE A CAL LO
AND CAL HI FIRST?
YES

Figure 42: Two point calibration screen

W-4510 Weigh Scale Technical Reference Manual
NO