Input Channel Setup; Frequency Input Channel #1, Primary Sensor Signal (Frequency Only); 521 Frequency Input Channel #2, Tachometer/Speed Input; Figure 31: Input Channel Setup Screens - Ohmart Vega W-4510 Technical Reference Hardware Manual

Weigh scale with gen 2000 electronics model / with frequency output
Table of Contents

Advertisement

Setup

Input channel setup

The screens at 520 and 521 give direct access to the parameters controlling the two input
channels. There is no need to access these screens for basic gauge operation. The
calibration functions automatically update these values as necessary.
Note: Personnel that are unfamiliar with the system should not access these screens.
Changing values at these address locations can adversely effect the operation of the gauge.
Each of the screens (520 and 521) hold 50 items numbered 0 to 49. Please refer to the
tables in Appendix I, page 130 for a complete listing of each item. Refer to Figure 24 for
jumper configurations for the frequency input Channel #1.
520 FREQIN CHAN#1
ITEM
000
CURRENT VAL
000000
520 Frequency input channel #1, primary sensor signal (frequency
only)
This memory block processes frequency input parameters such as:
Signal type (direct/inverse)
Source decay counts
Raw, filtered, normalized, and calibrated counts
Gain factors

521 Frequency input Channel #2, tachometer/speed input

This memory block processes frequency input signals from Channel #2 similar to Channel
#1. Channel #2 input is limited to optional equipment, such as, the tachometer.
It is possible, in limited application needing two sensors, for the raw count signal at Channel
#2 to come from a secondary sensor. The counts add to those coming into Channel #1 from
the primary sensor to obtain a total raw count, before signal processing. You cannot use this
feature to process two separate measurements. Refer to Figure 24 for jumper configurations
for the Frequency input Channel #2.
Tachometer
jumper configuration for frequency
Speed input
jumper configuration for 0–10V or 4–20mA input
64
NEXT
000
NEW VAL
000000

Figure 31: Input channel setup screens

W-4510 Weigh Scale Technical Reference Manual
521 FREQIN CHAN#2
ITEM
000
NEXT
CURRENT VAL
NEW VAL
000000
000000
000

Advertisement

Table of Contents
loading

Table of Contents