Download Print this page

Abbott CELL-DYN 3000 Troubleshooting Manual page 41

Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
IP - 10 Alarm 4 (d) Mixing Home And Limit Sensors, 1 of 1
S T AR T
- P o w e r S a m p le L o a d e r O F F
- V e rify G o o d E le c tric a l
C o n n ec tio n a t B o th H o m e
a n d L im it S e n s o rs , a n d a t
C P U M a in B d . C N 5
- C le a n B o th S e n s o rs
[NON- CE MARK]
[CE MARK]
Pin 10
Mixing Limit
Pin 4
Mixing Home
TROUBLESHOOTINGIP - 10 Alarm 4 (d) Mixing Home And Limit Sensors, 1 of 1
1
Pin 4 Mixing Limit
Pin 2 Mixing Home
TP
SL-131B
Search Book TOC Go Back
P ro b le m
C orre c te d
?
N o
M ixing (Z 2 )
H o m e S e n s o r a n d
L im it S en s o r =
+ 5 .0 V B loc k e d , 0 .0 V
U n b loc k e d
(S e e F ig u re )
?
Y e s
R e p la c e C P U
M a in B d .
(P L-8 .1
;
S e e
In te rre la tio n sh ip
M a trix -
P g . 3-4
)
9140234B-June 1997
Y e s
P e rfo rm T o ta l C a ll
R e p la c e D e fe c tiv e
S e n s o r
(P L -4 .1 1
In te rre la tio n s h ip
M a trix -
N o
M ix in g
C o n ta c t
S e n s o r
M ix in g L im it
S e n s o r
; S e e
P g . 3 -4
)
M ix in g
H o m e
S e n s o r
S L - 1 2 1 D
1 - 32

Advertisement

loading

This manual is also suitable for: