Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
4. Align Welded Platen:
a. Welded Platen walls cannot be straightened. If walls are not square, replace Platen.
b. It may be necessary to adjust Rack Guide and/ or Beam Block for Tool to fit properly.
To Adjust Rack Guide:
-
Loosen (2) screws attaching Rack Guide to Platen.
-
Insert Rack Guide Alignment Tool, adjust Rack Guide, and tighten hardware.
To Adjust Beam Block:
-
Go to
"VP - 3.5 Rack Clearance Adjustment" (Page
5. Go to
"VP - 3.4 Rack Clearance Check" (Page
3
VERIFICATION PROCEDURESVP - 3.3 Platen Wall Adjustment
Search Book TOC Go Back
3-28).
3-26).
9140234B-June 1997
3 - 24