CELL-DYN
3/35 Sample Loader TS Guide
VP - 3.3 Platen Wall Adjustment
Purpose: Perform this procedure to adjust Platen Walls, Rack Guides, and Beam Block.
1. Determine if Platen is welded or non-welded. Non-welded Platen has 14 wall screws on top of
Platen.
2. If Platen is non-welded, go to Step 3. If Platen is welded, go to Step 4.
3. Align Non-Welded Platen:
a. Loosen Platen Wall screws (1/2 turn) at right end of Platen.
b. Insert Rack Guide Alignment Tool at Right end (1) of Platen. Ensure Walls are tight against
Tool then tighten screws.
3
VERIFICATION PROCEDURESVP - 3.3 Platen Wall Adjustment
Search Book TOC Go Back
Special Tools
DESCRIPTION
Rack Guide
Alignment Tool
QTY
PN
1
9420084
9140234B-June 1997