3/35 Sample Loader TS Guide
IP - 5 No Lights, Some Other Functions Work, 1 of 3
- E n s u re P o w e r C o rd P ro p e rly
C o n n e c ted a n d S a m p le L o a d er
P o w e r O N (1 )
- S e a t C N 1 o n C P U M a in B d
- S e a t C a ble fro m C P U M a in B d
C N 2 to K e y b o a rd C N 2
- C h e c k + 5V to K e y b o a rd , S e e
fig u re o n th is p a g e
K e y b o a rd C N 2
P in 1 7 (+ ) to G N D
= + 5 V ±0 .2
TROUBLESHOOTINGIP - 5 No Lights, Some Other Functions Work, 1 of 3
1
S T AR T
?
N o
1
Search Book TOC Go Back
[CE MARK ]
Pin 17 to GND = +5V




Back of
New
Keyboard
Y e s
R e p lac e K e y b o a rd
In te rrela tio n s h ip M a trix -
9140234B-June 1997
[NON- CE MARK]
Pin 17 to GND = +5V
CN2




(P L -9 .3
; S e e
P g. 3 -4
)
SL-116B