Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
5. Verify Contact Sensor Board Position.
E and F Loaders
[CE MARK] Top of Mixing Paddle must be .040-.060 in. (1-1.5 mm) below bottom of Sensor
when Mixing Head is in Home Position (Z
[NON-CE MARK] E Loader - Top of Mixing Paddle must be flush with bottom of Sensor
(when Mixing Assembly is home,
3
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
[CE MARK]
.040 - .060 in
(1.0 - 1.5 mm)
E- Loader
Figure 3-46
Figure 3-48,
Search Book TOC Go Back
). See
Figure 3-46
SL-269
Page-3-95).
9140234B-June 1997
and
Figure 3-47,
Page-3-94.