CELL-DYN
3/35 Sample Loader TS Guide
VP - 3.8 Rack to Laser Shield Check
Purpose: Perform this procedure to ensure Platen is aligned (side to side) with the Tower.
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
c. Ensure BAR (1/0) Switch is set to 1
2. Ensure Needle is in Home position.
3. Ensure Laser Shield hardware is tight.
4. Press (
X) key to move X-Axis Solenoids to Left Stop.
5. Insert Rack Guide Alignment Tool in Right Side of Platen.
6. Place Rack in Tool cutout (Under Tower)
7. Both Laser Shield Cutouts must line up exactly with tube positions 3 and 5.
8. Press MAN to energize tube pusher solenoid.
9. Ensure tube pusher arm(s) do not hit rack.
3
VERIFICATION PROCEDURESVP - 3.8 Rack to Laser Shield Check
Search Book TOC Go Back
Special Tools
DESCRIPTION
Rack Guide
Alignment Tool
(Figure
QTY
PN
1
9420084
3-17).
9140234B-June 1997
3 - 39