®
CELL-DYN
3/35 Sample Loader TS Guide
N) key to move Needle to Home position (UP).
7. Press (
↑
8. Ensure aspiration hole in needle is centered in waste cavity line and tip of needle is flush with
bottom of Wash Block.
NOTE:Needle must not extend below bottom of Wash Block nor recess more than 0.040" (1mm).
Hole Centered on
Waste Drain Cavity
3
VERIFICATION PROCEDURESVP - 8.1 Needle Sensors Check/Adjustment
Search Book TOC Go Back
Limit (Bottom):
Tip 1/2" (~1 cm) Above Platen
Figure 3-65
E LOADER WITH
BROWN WASH BLOCK
Wash Block
Home: Tip Flush with
Bottom of Wash Block or
Recessed Up to .040 in.
SL-274
9140234B-June 1997
3 - 123