Download Print this page

Abbott CELL-DYN 3000 Troubleshooting Manual page 168

Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
VP - 3.10 X-Axis Solenoids to Platen Channel Check
Purpose: Perform this procedure to verify Platen Channels are parallel to X-Axis Guide Rods.
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Press (
X
keys, move the solenoids along their full length of travel.
→)
3. Visually confirm the following:
The solenoids complete the movement without touching the Platen.
The gap between each solenoid and the Platen Channel is approximately the same all along
the length of travel.
4. IF Check is OK, go to
5. IF Check Fails, go to
3
VERIFICATION PROCEDURESVP - 3.10 X-Axis Solenoids to Platen Channel Check
Search Book TOC Go Back
Special Tools
DESCRIPTION
None
"VP - 3.12 Platen to Laser Shield Check" (Page
"VP - 3.11 X-Axis Solenoids to Platen Channel Adjustment" (Page
QTY
PN
9140234B-June 1997
3-47).
3-45).
3 - 43

Advertisement

loading

This manual is also suitable for:

Cell-dyn 3500