CELL-DYN
3/35 Sample Loader TS Guide
1.1 OVERVIEW
This chapter contains the following information
for Type E, F, & G Sample Loaders:
1.2 Error Messages
1.3 Isolation Procedures (IPs)
Troubleshooting Approach
Troubleshooting is required whenever an
Observed Error or a Displayed Error occurs.
Sample Loader Displayed Errors can be
viewed on the Bulletin Line of the Data Station
screen. (The Bulletin Line is down about 3/4
way on the screen.)
When a malfunction occurs, follow these
recommended procedures:
1. Decontaminate the Sample Loader and the
Analyzer according to General Biosafety
Procedure.
2. Use this Troubleshooting Guide as follows:
a. If possible, begin by verifying the
reported error. Perform the exact pro-
cedure that produced the error condi-
tion.
TROUBLESHOOTING
1
Search Book TOC Go Back
b. When an error occurs, either observed
or displayed, look up the error in Sec-
tion 1.2, Error Messages table.
c. Perform the Corrective Action.
d. If directed to an IP, flip pages while look-
ing on edge of sheets for the desired IP
Number.
e. Turn to page 1 of x, and begin at the
START box.
f. Follow the directions in the IP.
g. If directed to perform a Verification Pro-
cedure (VP), flip pages while looking on
edge of sheets for the desired VP Num-
ber.
h. If a part is adjusted or replaced, go to
Section 3.2, Interrelationship Matrix.
i. Find part in left column.
j. Perform VPs in order given (i.e. 1, 2, 3,
4, etc.)
3. Perform Total Call.
1.1 OVERVIEW
9140234B-June 1997
--- END ---