CELL-DYN
3/35 Sample Loader TS Guide
VP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
Purpose: Perform this procedure to verify and adjust Vent Needle Alignment, F Loader Only,
3-70.
1. Perform Steps 1, 2, and 3 of
2. Ensure proper Needle is installed and Needle Port is centered on Wash Port
3
VERIFICATION PROCEDURESVP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
Search Book TOC Go Back
Special Tools
DESCRIPTION
Needle Alignment
Tool
Rack Guide
Alignment Tool
"VP - 9.1 Sample Wash Block Check/Adjustment" (Page
QTY
PN
1
9420081
1
9420084
9140234B-June 1997
Figure
3-125).
(Figure
3-69).