Abbott CELL-DYN 3000 Troubleshooting Manual page 160

Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
c. Check distance between Tool and solenoids using a Feeler Gauge. Distances should be the
same (±.005 in., 0.127mm).
d. If Check is OK, Install Platen then go to
e. If Check Fails, go to
.
3
VERIFICATION PROCEDURESVP - 3.6 X-Axis Solenoid to Guide Rod Check
Search Book TOC Go Back
"VP - 3.8 Rack to Laser Shield Check" (Page
"VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment" (Page
Alignment Tool
Solenoid
Figure 3-14 Solenoid Parallelism
Same
Distance
Same
Distance
SL-235
--- END ---
9140234B-June 1997
3-39).
3-36).
3 - 35