Download Print this page

Abbott CELL-DYN 3000 Troubleshooting Manual page 249

Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
9. Press (N
) key to move Needle to the maximum DOWN position.
10.Verify distance between needle tip and Platen floor
11.If check fails, adjust sensors.
3
VERIFICATION PROCEDURESVP - 8.1 Needle Sensors Check/Adjustment
G Limit=
1.0"-1.1"
(25.4mm-27.9mm)
Floor of Platen
Figure 3-66 Sample Needle Home and Limit Clearances
Search Book TOC Go Back
F and G
LOADER
F and G Home =
Flush or Recessed
to .040 in.
F
Limit = 0.5" (12.7mm)
(Figure
--- END ---
9140234B-June 1997
SL-109B
3-66).