Abbott CELL-DYN 3000 Troubleshooting Manual page 221

Sample loader
Hide thumbs Also See for CELL-DYN 3000:

Advertisement

®
CELL-DYN
3/35 Sample Loader TS Guide
6. Verify Contact Sensor clearance.
[ CE MARK ] Clearance Check is NOT required.
[NON- CE MARK ] Mixing Paddle must clear Sensor by .010-015in. (0.25-0.38mm.); .015in.
(0.38mm) preferred
3
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
Flush With
Bottom of Sensor
G Loader
(Figure 3-50,
Page-3-97).
Search Book TOC Go Back
All G Loaders
Figure 3-49
9140234B-June 1997
SL-271
3 - 96