CELL-DYN
3/35 Sample Loader TS Guide
VP - 3.12 Platen to Laser Shield Check
Purpose: Perform this procedure to verify Platen is aligned (front to back) with the Tower.
1. Ensure Laser Shield hardware is tight.
2. Verify Laser Shield is slightly recessed from Platen Wall. Check Left and Right end of Laser
Shield
(Figure
3-21).
IF Check is OK, go to
IF Check Fails, go to
3
VERIFICATION PROCEDURESVP - 3.12 Platen to Laser Shield Check
Search Book TOC Go Back
Special Tools
DESCRIPTION
None
"VP - 3.14 Pusher Arm to Platen Check" (Page
"VP - 3.13 Platen to Laser Shield Adjustment" (Page
QTY
PN
9140234B-June 1997
3-51).
3-49).
3 - 47