3/35 Sample Loader TS Guide
VP - 4.1 Rack Sensor Check
Purpose: Perform this procedure to verify Rack Sensor Operation.
All Loaders
1. Remove all Racks from Sample Loader.
2. Clean all Racks according to Operator's Manual.
3. Clean Rack Sensors A through E with cotton tip applicator and water
4. Ensure sensor surface does not protrude beyond platen wall. Sensor face must be flush or
recessed not more than 1/2 thickness of platen wall
go to
"VP - 4.2 Rack Sensor Adjustment" (Page
5. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
c. Set BAR (1/0) Switch to 1.
3
VERIFICATION PROCEDURESVP - 4.1 Rack Sensor Check
Search Book TOC Go Back
Special Tools
DESCRIPTION
Feeler Gauge Set
QTY
PN
1
1-43627-01
(Figure
3-27). If sensor position is incorrect,
3-63).
9140234B-June 1997
(Figure
3-28).
3 - 59