CELL-DYN
3/35 Sample Loader TS Guide
9140234B-June 1997
6. Ensure tube positioning arm(s) does not drag on top of platen. If necessary, loosen solenoid
bracket screws and adjust.
Only [OLD] F Loader with Clamp - (Others not Adjustable)
a. To raise arms slightly, loosen screws on solenoid bracket
(Figure
3-74).
b. Push bracket up to give clearance.
3
VERIFICATION PROCEDURESVP - 10.2 Tube Positioning Arm Alignment
3 - 137