Installing/Exchanging The Flash Bulb - Leica IP C Instructions For Use Manual

Automated printing system for tissue cassettes
Hide thumbs Also See for IP C:
Table of Contents

Advertisement

4
Instrument Setup
6
1
Fig. 16
4.6

Installing/exchanging the flash bulb

Removing the old flash bulb
Warning
Switch the instrument off and unplug it from power supply. Allow the flash bulb to cool off before
removing it. Do not handle the flash bulb with bare hands. Use a glove or tissue.
1. Open the lid
2. Remove screw
(→ "Fig.
17-2").
3. Swing the reflector
32
5
2
(→ "Fig. 15-6")
to gain access to the reflector
(→ "Fig. 17-3")
(use screwdriver supplied as part of tool set). Mind the washer
(→ "Fig. 17-1")
upwards.
2
3
4
(→ "Fig.
17-1").