7
Troubleshooting
Display
code
Cassette not correctly clamped in cassette
carrier.
Cassette did leave the feeding chute but
did not reach the cassette carrier.
Cassette did not leave the cassette carrier
or was still located in the cassette carrier
during initialization.
The ink print head is getting too hot.
Ambient temperature too high or
electronics defective.
No or incorrect voltage at ink print head.
• Cover
• Flash standby state not reached within
No flash or flash duration too short.
Flash bulb dirty or defective.
Unload station mechanically blocked.
Heater fan not running or running too
slowly.
No cassette is in the drying module for
flashing. Cassette left the cassette carrier,
but did not reach the drying module.
Cassette not successfully ejected from
drying module.
76
Source of error
(→ "Fig. 2-5")
not closed properly.
Safety switch triggered.
the prescribed time. Charging electronics
defective.
Troubleshooting solution
Remove the cassette from the cassette
carrier.
Remove the cassette from the cassette
carrier.
Switch instrument off and let it cool off.
The instrument remains disabled until
the print head has cooled down to a
temperature value within the permissible
range.
Check ambient temperature.
Call Leica Technical Service.
• Ensure that the cover is not blocked,
e.g. by the manual unload station. Close
cover completely.
• Call Leica Technical Service.
Check whether flash bulb works – to do
so, observe the stray light on the cover.
Never open the cover to check whether a
flash is triggered!
No flash --> install a new flash bulb
(→ p. 32 – 4.6 Installing/exchanging the
flash
bulb).
Remove the cause of the blockage.
Call Leica Technical Service.
Remove cassette.
Remove cassette.