Fig. 54
4. Moisten one of the foam swabs supplied with the instrument
Be sure not to use too much alcohol – no alcohol may drip into the instrument.
Warning
Never use acetone or xylene! Only use 95 % or 100 % alcohol for cleaning purposes.
5. Carefully insert the swab into the gap under the print head
upwards (on the print head) and move the foam swab back and forth (approx. 10 times). This
procedure removes dried ink residues.
Warning
Never rotate the swab – this can damage the nozzle plate of the print head.
Fig. 55
6. Also clean the
sealing lip
7. Check the sealing lip for damage. Replace the
Leica IP C
1
replacement plate
(→ "Fig. 56-1")
(→ "Fig. 56-2")
must be completely free of ink residue.
Cleaning and Maintenance
(→ "Fig. 55-1")
(→ "Fig.
55"). Apply light pressure
and sealing lip with (clean) alcohol. The
replacement plate
if the sealing lip is damaged.
6
with some alcohol.
65